15-00044
Revision 5
April 2008
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 21
Set switch 7&8 to the OFF position only in case the meter is being stored.
Important Note:
For operational/configuration settings ensure dipswitches 1, 2, and 3 (dipswitch 3 is for
transmitting data) are in the ON position. For field use ensure dip switches1, 2, 3, 7 and
8 are turned ON (7 and8 are for battery backup).
4.2.5 LED position and function
LED Color Function
Ext. power Red External power supply > 7.5v – led on
External power supply < 7.5v – led blinks
No external power supply & no Battery – LED off
Tx. Green For each transmission sent, the led blinks
Battery Yellow Battery connected & Battery voltage > 7.5v – LED is on
Battery disconnected or Battery voltage < 7.5v – LED blinks
No external power supply & no Battery – LED is off
Rate Yellow Blinks according the radiation trend