15-00043
Revision 3
September 2005
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 8
The PAM-TRX is designed to transmit the TTL RS-232 data packet from the dosimeter when the
DMC-2000 is configured with transmission ON, 4800 BAUD, Triggered. The unit is compatible
with all the DMC family gamma dosimeters data formats (DMC2000S, X and SOR/R).
Preparation for use:
2.1.2.1 To open the back case: remove the locking screw and slide off the battery cover. Remove
the battery (if installed) and then slide off the case back.
2.1.2.2 The dosimeter must be installed with the display towards the top window face down and
dosimeter clip removed. Once inserted, replace the back cover by sliding it up, ensuring
the dosimeter is properly seated and aligned with the ISO connector.
2.1.2.3 Ensure dipswitch settings are configured with all the desired options correctly selected.
See Section 2.1.3 for detailed configuration.
2.1.2.4 Install a new ‘AA’ 1.5 V alkaline battery. Replace the battery cover and secure with the
locking screw. Do not overtighten the locking screw
. MGPI recommends using only
Duracell or Energizer brand batteries for best performance.
>> See Figure 1 for PAM-TRX part identification.
2.1.2.5 Turn the dosimeter ON. Note one single flash from the green led will indicate that the
battery, dosimeter and the radio are functioning properly. After one or two short alarm
cycles (while the low battery detector charges up), the PAM-TRX audibly chirp. At this
point, the green LED will flash at an interval defined by the radio configuration, typically
four (4) seconds. The radio is normally in a “sleep mode” to reduce power consumption.
When the radio “wakes up,” it turns on the PAM-TRX power supply. The alarm circuits
are enabled at all times. Pressing the alarm test button or running a magnet by the case side
(internal reed switch and if the option is enabled) will turn ON the power supply and trigger
all enabled alarms.
2.1.2.6 If the dosimeter alarms (dose or dose rate alarm), the enabled alarm functions in the
PAM-TRX will be triggered (buzzer, vibrator, LED).
2.1.2.7 The alarm can be tested at any time by pressing a test push button accessible through a
small hole in the back of the case or by proximity to a magnet by the side of the case if
this function is enabled (Figure 2 below).
Figure 2: Alarm Test Button Access
Zobrazit stránku 7